Featured Product
LAUDA Alpha Heating
& Cooling Thermostats
Heating and cooling
thermostats for cost
effective thermostating
in the laboratory at
temperatures from
-25 up to 85 °C.
Guaranteed to be in stock for immediate delivery!
Process circulation chillers for industrial applications with cooling capacities up to 265 kW at working temperatures from -5 up to 25 °C in ambient conditions from -15 up to 50 °C
LAUDA Ultracool chillers provide reliable temperature control and ensure secure processing. The chillers of the UC series are available as Ultracool Standard (ST) and Ultracool Superplus (SP). All chillers are already equipped with an antifreeze protection thermostat to prevent freezing of the heat exchanger.
Integrated pressure switches protect the circuit against pressure levels that drift too high or too low. Housings made from galvanized steel and externally coated with epoxy resin protects against corrosion even under aggressive ambient conditions. Most of the models are suitable for outdoor installation. The Ultracool Standard units provide the necessary cooling capacity for customers who already have a pump and a water reservoir installed. The Superplus models are plug & operate systems equipped with a cold water tank, a centrifugal pump and an internal bypass
| The Ultracool advantages | Your benefits | |
![]() |
|
|
![]() |
|
|
![]() |
|
|
![]() |
|
|
![]() |
|
|
The three UC Mini chillers are available with cooling capacities from 1.9 up to 4.7 kW. The models work with a reciprocating compressor and a centrifugal pump. The pump is noise reduced and works with a very flat flow rate/pressure characteristic. This allows the user to easily adjust the water flow without jeopardizing the pressure.
By using R 134a as refrigerant, the pressure inside the refrigeration system is rather low and therefore the units can work in ambient temperatures up to 50 °C.
Six UC Midi types have cooling capacities ranging from 7.1 up to 26.3 kW. The models work with a reciprocting or scroll compressor and a centrifugal pump. The noise-reduced pump allows customers to easily adjust the water flow. The use of R 134a as refrigerant ensures very low working pressures inside the refrigeration system and operation in ambient temperatures up to 50 °C . The intergrated motor fan speed regulator allows operation in ambient conditions up to -15 °C and reduces the noise level additionally.
The models UC-0060 to UC-0240 are also available as UC laser models with pre-configured options included.
Within the Ultracool Maxi range, there are four chillers with cooling capacities from 34.1 up to 64.4 kW which work with one single refrigeration circuit. The five models from 87.9 up to 265 kW of cooling capacities work with two independent refrigeration circuits to provide a backup security. All Maxi models are equipped with scroll compressors, work with R 407C as refrigerant and are suitable for outdoor installation. The internal pipes for the water circuit are made of PP-R (polypropylene random copolymer) and are thermowelded. Besides the main advantage that these connections are absolutely leak free, PP-R is corrosion and frost proof, allow for smaller pressure drops and are long lasting.
The models UC-0300 to UC-0650 are also available as UC laser models with pre-configured options included.
Select up to 3 products to compare: Compare Products
| Technical features | UC-0020 ST | UC-0030 ST | UC-0040 ST | |
| Select to compare: | ||||
| Working temperature range | °C | -5…25 | -5…25 | -5…25 |
| Ambient temperature range | °C | 5…50 | 5…50 | 5…50 |
| Cooling output* | kW | 1.9 | 3.5 | 4.7 |
| Cat. No. 230 V; 50 Hz | E6813102 | E6813104 | E6813106 | |
| Cat. No. 220 V; 60 Hz - | E6813302 | E6813304 | E6813306 | |
| Technical features | UC-0020 SP | UC-0030 SP | UC-0040 SP | |
| Select to compare: | ||||
| Working temperature range | °C | -5…25 | -5…25 | -5…25 |
| Ambient temperature range | °C | 5…50 | 5…50 | 5…50 |
| Cooling output* | kW | 1.9 | 3.5 | 4.7 |
| Pump pressure nominal | bar | 3.5 | 3.3 | 3.1 |
| Pump flow nominal | L/min | 5.6 | 10.3 | 13.8 |
| Volume water tank | L | 35 | 35 | 35 |
| Cat. No. 230 V; 50 Hz | E6813002 | E6813004 | E6813006 | |
| Cat. No. 220 V; 60 Hz | E6813202 | E6813004 | E6813206 | |
* at 10 °C water outlet temperature and 25 °C ambient temperature, for 50 Hz versions
| Technical features | UC-0060 ST | UC-0080 ST | UC-0100 ST | UC-0140 ST | UC-0180 ST | UC-0240 ST | |
| Select to compare: | |||||||
| Working temperature range | °C | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 |
| Ambient temperature range | °C | -15…50 | -15…50 | -15…50 | -15…50 | -15…50 | -15…50 |
| Cooling output* | kW | 7.1 | 9.4 | 11.7 | 14.5 | 22.0 | 26.3 |
| Cat. No. 400-440 V; 3/PE; 50 Hz | E6813108 | E6813110 | E6813112 | E6813114 | E6813116 | E6813118 | |
| Cat. No. 440-480 V; 3/PE; 60 Hz | E6813308 | E6813310 | E6813312 | E6813314 | E6813316 | E6813318 | |
| Technical features | UC-0060 SP | UC-0080 SP | UC-0100 SP | UC-0140 SP | UC-0180 SP | UC-0240 SP | |
| Select to compare: | |||||||
| Working temperature range | °C | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 |
| Ambient temperature range | °C | -15…50 | -15…50 | -15…50 | -15…50 | -15…50 | -15…50 |
| Cooling output* | kW | 7.1 | 9.4 | 11.7 | 14.5 | 22.0 | 26.3 |
| Pump pressure nominal | bar | 4.0 | 4.0 | 3.9 | 3.7 | 3.2 | 2.7 |
| Pump flow nominal | L/min | 20.1 | 26.6 | 33.6 | 43.8 | 62.6 | 84.1 |
| Volume water tank | L | 100 | 100 | 100 | 100 | 100 | 100 |
| Cat. No. 400-440 V; 3/PE; 50 Hz | E6813008 | E6813010 | E6813012 | E6813014 | E6813016 | E6813018 | |
| Cat. No. 440-480 V; 3/PE; 60 Hz | E6813208 | E6813210 | E6813212 | E6813214 | E6813216 | E6813218 | |
* at 10 °C water outlet temperature and 25 °C ambient temperature, for 50 Hz versions
| Technical features | UC-0300 ST | UC-0400 ST | UC-0500 ST | UC-0650 ST | UC-0800 ST | UC-1000 ST | UC-1350 ST | UC-1700 ST | UC-2400 ST | |
| Select to compare: | ||||||||||
| Working temp. range |
°C | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 |
| Ambient temp. range |
°C | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 |
| Cooling output* | kW | 34.1 | 43.3 | 48.7 | 64.4 | 87.9 | 106.4 | 139.2 | 175.7 | 265.0 |
| Cat. No. 400-440 V; 3/PE; 50 Hz |
E6813120 | E6813122 | E6813124 | E6813126 | E6812128 | E6812130 | E6812132 | E6812134 | E6812136 | |
| Cat. No. 440-480 V; 3/PE; 60 Hz |
E6813320 | E6813322 | E6813324 | E6813326 | E6812328 | E6812330 | E6812332 | E6812334 | E6812336 | |
| Technical features | UC-0300 SP | UC-0400 SP | UC-0500 SP | UC-0650 SP | UC-0800 SP | UC-1000 SP | UC-1350 SP | UC-1700 SP | UC-2400 SP | |
| Select to compare: | ||||||||||
| Working temp. range | °C | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 | -5…25 |
| Ambient temp. range | °C | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 | -15…45 |
| Cooling output* | kW | 34.1 | 43.3 | 48.7 | 64.4 | 87.9 | 106.4 | 139.2 | 175.7 | 265.0 |
| Pump pressure nominal | bar | 3.9 | 3.6 | 3.3 | 3.7 | 3.4 | 3.3 | 4.3 | 3.6 | 3.8 |
| Pump flow nominal | L/min | 98.0 | 123.0 | 150.0 | 196.0 | 247.0 | 299.0 | 392.0 | 494.0 | 733.0 |
| Volume water tank | L | 200 | 200 | 200 | 300 | 300 | 500 | 500 | 500 | 500 |
| Cat. No. 400-440 V; 3/PE; 50 Hz |
E6813020 | E6813022 | E6813024 | E6813026 | E6812028 | E6812030 | E6812032 | E6812034 | E6812036 | |
| Cat. No. 440-480 V; 3/PE; 60 Hz |
E6813220 | E6813222 | E6813224 | E6813226 | E6812228 | E6812230 | E6812232 | E6812234 | E6812236 | |
* at 10 °C water outlet temperature and 25 °C ambient temperature , for 50 Hz versions